Immunofluorescence Assay CK-MB/cTnI/Myo Combo Fast Test Kit

Short Description:

Diagnostic Kit for Isoenzyme MB of Creatine Kinase

Methodology: Fluorescence Immunochromatographic Assay


  • Testing time: 15 minutes
  • Storage temperature: 2°C-30°C
  • Valid Time: 24 months
  • Accurancy: More than 99%
  • Specification: 25 test/box
  • Product Detail

    Product Tags

    1. INTEDED USE

    This kit is applicable to in vitro quantitative detection of concentrations of myocardial injury markers of cardiac troponin I, isoenzyme MB of creatine kinasein and myoglobin in human serum/plasma/whole blood sample, and it’s suitable for auxiliary diagnosis of myocardial infarction. This kit only provides test results of cardiac troponin I, isoenzyme MB of creatine kinasein and myoglobin, and results obtained shall be used in combination with other clinical information for analysis.

    2. PRODUCT SPECIFICATION

    Model No. CK-MB/CTNI/MYO 
    Methodology Fluorescence Immunochromatographic Assay
    Sample Type serum/Plasma/Whole Blood 
    Time to Result 10-15mins
    Storage 2~30 ℃/36~86℉
    Shelf Life 24 months
    Certificate ISO13485,CE,MHRA
    MAIN KIT COMPONENTS

    * Test device

    * Sample diluents

    * Instructions for Use

    PACKING

    * 25 test /kit

    * Aluminum foil bag labeling

    * shrink wrap

    3.TEST METHOD

    4.CLINICAL PERFORMANCE

    Clinical performance of this product’s assessed through collection of 150 cases of clinical samples. 

    a) In case of cTnI item, corresponding marketed kit of chemiluminescence assay’s used as reference reagent, detection results have been compared and their comparability has been studied through linear regression, and correlation coefficients of the two assays are Y=0.975X+0.074 and R=0.9854 respectively;

    b) In case of CK-MB item, corresponding marketed kit of electrochemiluminescence assays used as reference reagent, detection results have been compared and their comparability has been studied through linear regression, and correlation coefficients of the two assays are Y=0.915X+0.242 and R=0.9885 respectively.

    c) In case of MYO item, corresponding marketed kit of time-resolved fluor immunoassays used as reference reagent, detection results have been compared and their comparability has been studied through linear regression, and correlation coefficients of the two assays are y=0.989x+2.759 and R=0.9897 respectively.

    QQ图片20231025170447
    QQ图片20231025170447
    QQ图片20231025170447

    5. CERTIFICATE

    * ISO System Certificate

    * CE Certificate

    * EU Registration

    * UCKA MHRA Registration