Helicobacter Pylori antibdy Hp-Ab uncut sheet

Short Description:

Helicobacter Pylori antibody Hp-Ab uncut sheet

Methodology: Colloidal Gold


  • Testing time: 15 minutes
  • Storage temperature: 2°C-30°C
  • Valid Time: 24 months
  • Accurancy: More than 99%
  • Specification: 25 test/box
  • Product Detail

    Product Tags

    1. INTEDED USE

    This kit is applicable to in vitro qualitative detection of antibody to H.pylori (HP) in human whole blood, serum or plasma sample, which’s suitable for auxiliary diagnosis of HP infection. This kit only provides test results of antibody to H.pylori (HP), and results obtained shall be used in combination with other clinical information for analysis.

     

    2. PRODUCT SPECIFICATION

    Model No. HP-AB Uncut sheet
    Methodology Colloidal Gold
    Sample Type Serum,plasma,whole blood
    Time to Result 10-15mins
    Storage 2~30 ℃/36~86℉
    Shelf Life 24 months
    Certificate ISO13485, CE Certificate, UCKA MHRA Certificate
    uncut sheet
    uncut sheet

    Feature

      High sensitive

    •  result reading in 15-20 minutes

    •  Easy operation

    •  High Accuracy

    PACKING

    * 20 bags /CTN

    * Aluminum foil bag labeling

    * shrink wrap

     

    3.TEST METHOD

    1
    Remove test device from aluminum foil pouch, lie it on a horizontal workbench, and do a good job in sample marking.
    2
    In case of serum and plasma sample, add 2 drops to the well, and then add 2 drops of sample diluent dropwise. In case of whole blood sample, add 3 drops to the well, and then add 2 drops of sample diluent dropwise.
    3
    Interpret result within 10-15 minutes, and detection result is invalid after 15 minutes (see detailed results in result interpretation).

     

    4.RESULT EVALUATION AND EXPLANATION

    微信图片_20241028154133

    5.CLINICAL PERFORMANCE

    Clinical performance of this product’s assessed through collection of 331 cases of clinical samples. Sample providers include patients with abnormal hp-ab index, patients with fecal occult blood and normal people. Marketed kit of colloidal gold methods used as reference reagent. The WIZ BIOTECH reagent test will be compared with the control reagent:

    Baysen Result of hp--ab
    Test result of Reference reagent Positive coincidence rate:
    98.92(95%C.I.96.16%~99.7%)
    Negative coincidence rate:
    100.00(95%C.I.97.42%~100.00%)
    Total coincidence rate:99.65% (99.4%C.I. 97.82%~99.83%)
    Positive  Negative  Total
    Positive 184 0 184
    Negative  2 145 147
    Total 186 145 331